The author would like to express appreciation to Robert Douglas for providing
analytical support during the validation. This support included preparation of standards, sample spikes, and operation
of the ARL 3560. In Addition, the effort that he put forth in evaluating the sample digestion procedure is discussed
in Section 8 of this Backup Data Report.
1. Introduction
The purpose of this work was to validate a simultaneous Inductively Coupled Plasma-Atomic Emission Spectrometer (ICP-AES)
- an Applied Research Laboratories (ARL) Model 3560 (Fisons, Sunland, CA) for use in the analysis of industrial
hygiene brazing and soldering fume samples. Previously, the OSHA Salt Lake Technical Center (SLTC) had been analyzing
solder and brazing fume samples using a Jobin Yvon Model JY 70 ICP (Instruments SA, Edison, NJ)
(11.1).
1.1 Evaluation of the Analytical Method for Measuring Occupational
Exposures during Soldering and Brazing Operations
The main concern of this evaluation is to determine whether this instrument is able to detect, correctly
identify, accurately and precisely measure exposures near the respective PELs. The general procedure for the
analysis of samples by an ARL ICP has been previously described (11.2)
and evaluated (11.3). This
validation examines the analytical response of an ARL spectrometer to eight elements commonly found in soldering and
brazing operations. The eight elements (Ag, Be, Cd, Cu, Pb, Sb, Sn, and Zn) have been validated using concentration
levels of approximately 0.5, 1, and 2 times the Transitional or Final Rule Permissible Exposure Limits (PELs) (11.4)
as Time Weighted Averages (TWAs).
[Note: After the evaluation, the regulation for exposure to cadmium was changed to 0.005 mg/m³ as a
TWA. Please see the Code of Federal Regulations (29 CFR 1910.1027 or 29 CFR 1926.63) for further information.]
In addition, this method is sensitive enough to determine the Short-Term Exposure Limit (STEL) and Ceiling
concentrations for Be. This procedure does not differentiate between dusts and fumes.
Spiked samples were used to evaluate the method. Each spike was calculated to represent the amount of analyte
found at a concentration near its respective PEL when taking a 400- to 500-L sample. This air volume range is
approximately equivalent to taking 4-h samples when collected at a flow rate of 2 L/min.
1.2 Spectrometer Description
The ARL spectrometer includes a 1.0-m path length with 1,080 grooves/mm grating and a solid-state radio frequency
(RF) generator. Standard operating conditions for this instrument were developed through a series of benchmark tests
recommended for most analytical applications. Equipment and operating conditions used during this evaluation are
listed in Table 1. This
36-channel instrument was configured for the simultaneous determination of 35 elements, as listed within the line
library in Table 2. Two
channels are used for Fe; one wavelength (259.94 nm) is appropriate for dilute concentrations, while the other
(271.44 nm) is used for quantitating higher concentrations of Fe. The line library contains information for the line
array wavelengths and spectral orders as set up by the instrument manufacturer. The resolution at a specific
wavelength is dependent upon the grating and spectral-order of the line.
Spectral resolution is defined as full width at half maximum (FWHM), i.e., at half the height, the measured width
of the peak signal. Using the grating specified in
Table 1, the
resolution/spectral order relationship is described below:
Table of Order vs. Spectral Resolution
|
| Order |
Wavelength Range
(nm)* |
Resolution
(nm) |
SAMI Range
(motor steps) |
|
| 1st |
400-800 |
0.046 |
80 |
| 2nd |
235-400 |
0.023 |
| 3rd |
165-235 |
0.015** |
|
| * |
wavelength range is general - slight overlaps into the next order are possible |
| ** |
third order provides the best resolution |
|
The SAMI is a spectral scanning device that is further described in Section 3.1.
As shown above, a second-order peak has a resolution of 0.023 nm. A ±80 motor step scan of a second-order peak
would easily cover a range of 0.046 nm.
The analysis of additional elements beyond the eight validated allows for a more complete spectral interference
correction and provides for additional screening capability. A list of the elements screened and validated is
provided in Table 2 and
further displayed as calibration standards in
Table 3.
1.3 Standard and Reagent Purity
a) Standards:
Standards used during this study were prepared in a 32% HCl/4% HNO3 matrix except as noted in the
text. Single and multi-element standards were used to evaluate the instrument. Chemical compatibility of the
constituents was taken into account in order to avoid precipitation when preparing the multi-element standard
solutions.
Two multi-element standards and a reagent blank, as listed in
Table 3, were used for
instrument calibration of the validated elements. The reagent blank consisted of a 32% HCl/4% HNO3
mixture. A third standard solution (STD SOLN 3) is used to calibrate the screened elements.
Note: The concentration and combination of elements in each standard mixture was selected to minimize interferences
for the particular instrument and wavelengths used. Analysts using instruments with different wavelengths/elements
than stated may have to use alternate mixtures.
All standard solutions used for instrument calibration
(Table 3)
or spectral interference determinations were prepared by serial dilution.
b) Chemicals Used for Validation:
Sample Digestions and Dilutions
Nitric Acid (HNO3, 69.0 - 71.0% w/w) and Hydrochloric Acid (HCl, 36.5 - 38.0% w/w) ("Baker Analyzed"
Reagents, J.T. Baker Chemical Co., Phillipsburg, NJ).
Instrument Calibration
ICP Standard Solutions (1,000 µg/mL) ("Instra-Analyzed" Atomic Spectral Standards, J.T. Baker Chemical Co.)
Filter Spiking
Most solutions used for filter spiking were prepared from standards obtained from Inorganic Ventures, Inc.
(Brick, NJ). In addition, ICP standard solutions (10,000 µg/mL) from Specpure (Aesar/Johnson Matthey Inc., Seabrook,
NH) were used for any high concentration spikes.
Inter-element Corrections
Inorganic Venture and J.T. Baker standards were primarily used to determine inter-element
corrections.
1.4 Instrument Calibration
The polychromator was calibrated according to the manufacturer's procedures. The reagent blank and the
multi-element calibration solutions (Table
3) were matched to the sample matrix of 32% HCl/4% HNO3.
During the calibration process, linear regression coefficients were calculated by the Digital Equipment
Corporation (DEC) 11/53 computer from standard response readings. All calibrations were determined by first-order
regression only.
1.5 Soldering and Brazing Discussion
This validation primarily addresses air samples collected during soldering or brazing operations. The eight
elements most commonly found in these operations were: Ag, Be, Cd, Cu, Pb, Sb, Sn, and Zn. Other elements can be
added to this list. The capability for expanding the analysis to other elements is mainly dependent on laboratory
instrumentation, element solubility, and stability in the acid matrix used for digestion.
Applicable OSHA Permissible Exposure Limits (PELs) (11.4)
are listed in Table 4. This
Table lists both the Transitional and Final Rule PELs for substances which may be present in solder and brazing
operations. The Transitional and Final Rule Limits are identical with the exception of a Final Rule Ceiling Limit
for Be, a TWA for tin oxide, and additional STELs for zinc chloride and zinc oxide.
1.5.1 Welding, Soldering, and Brazing Applications (11.5,
11.6)
According to Kirk-Othmer (11.5),
welding, brazing, and soldering are metal-to-metal bonding processes. The temperature at which the joint is made is
the primary feature which differentiates between welding, soldering, or brazing. Solder alloys melt below 800°F
(427°C) and brazing alloys melt above 800°F.
a) Welding:
According to Kirk-Othmer (11.5),
in welding operations, similar components are fusion-bonded at or just below their melting points. The filler metal
is either puddled into relatively wide gaps, or the metal surfaces being joined are partially melted and bonded by
fusion or by a combination of puddling and fusion.
b) Brazing and soldering:
In most brazing and in almost all soldering operations, the components are molecularly bonded well below their
melting points. An exception is the brazing of aluminum and magnesium alloys. Generally, both soldering and brazing
involve the introduction of a non-ferrous filler material. While similar in concept, brazing and soldering are not
identical. The temperatures at which the alloys melt provide the primary difference between the two procedures.
The major brazing filler metals are copper, brass, bronze, and silver alloys. The filler metals are drawn into
closely fitted joints by capillary action and they bond and solidify without melting the components. Zinc and
cadmium can volatilize from zinc- and cadmium-containing brazing alloys during brazing.
1.5.2 Specific Components of Solders
The composition and use of some common solders is shown in
Table 5. A more detailed
description of some of the common solder alloys is given below:
a) Lead/tin:
Previously, most solder alloys were composed of combinations of tin and lead.
b) "Lead-free":
Recently, "lead-free" solders have become more prevalent. A sample of this type of solder was obtained and a
qualitative analysis by ICP and X-ray fluorescence (XRF) procedures at the SLTC confirmed this solder contained an
alloy of tin, silver, copper, and bismuth. The "lead-free" solders have been reported as possessing characteristics
similar to "50 Tin/50 Lead" solders, and a greater tensile and shear strength than those made with "50 Tin/50 Lead"
or "95 Tin/5 Antimony" solders. "Lead-free" solders have been substituted for lead solders in plumbing
applications, but have not been noticeably used in the electronics industry. This lack of use in the electronics
industry is apparently due to the high temperature needed to achieve the "lead free" solder's melting point.
c) Silver:
Many silver solders contain cadmium in varying amounts.
d) Antimony/tin:
These solders are also used as brazing alloys.
e) Cadmium/silver:
These are used with much higher temperatures and are suitable for use on copper and aluminum. This solder will
also produce a very high tensile strength.
f) Zinc:
Many of the alloyed solders may contain varying amounts of zinc.
g) Indium:
This is used for special applications, i.e., adhering glass to glass or glass to metal. The low vapor pressure
of these solders makes them useful for seals in vacuum systems.
h) Other metals:
Other trace contaminants present in base and filler metals include arsenic, chromium, bismuth, cobalt, nickel,
selenium, thallium, and vanadium.
1.5.3 Solder and Brazing Health Hazards (11.5,
11.6)
Solder's greatest danger to health lies in the presence of lead or cadmium in the solder alloy. Historically,
lead was present in large amounts in most solder alloys, and some special alloys contained cadmium.
In the past, the four most hazardous metals commonly found during soldering and brazing processes were lead,
cadmium, beryllium, and zinc. More recent solder formulations attempt to exclude most of these elements. Some of
the potential symptoms and hazards incurred from exposure to these and other elements are listed below (11.6)
and current PELs for validated elements are listed in
Table 4.
a) Lead:
Lead is used in the soldering process in the form of lead/tin and lead/silver filler metals. When heated, lead
oxide fumes are formed. Excessive exposure to lead oxide fumes can result in lead poisoning. Symptoms include loss
of appetite, indigestion, nausea, vomiting, constipation, headache, abdominal cramps, nervousness, and insomnia.
According to Kirk-Othmer (11.5),
lead is absorbed through the mucous membranes of the lung, stomach, or intestines and then enters the bloodstream.
b) Cadmium:
Cadmium is found in some silver and zinc solders, and in some base metals. When heated, cadmium oxide fumes can
be generated. Excessive exposure to these fumes can result in cadmium poisoning, symptoms of which include dry
cough, irritation of the throat and nasal passages, ulceration of the nose, tightness of chest, restlessness, and
renal damage. Cadmium is a suspected carcinogen and a higher incidence of prostate and lung cancers are noted among
workers in occupations that use cadmium in their processes.
c) Beryllium:
Beryllium is used in magnesium filler metals for furnace brazing and in some aluminum brazing filler metals.
While soldering, temperatures are normally too low to generate fumes from beryllium; however, the heat involved in
brazing can generate beryllium fumes, which are extremely hazardous. Short-term exposure to these fumes may result
in a chemical pneumonia. Long-term effects include shortness of breath, chronic cough, loss of weight, and fatigue.
Beryllium is a suspect human carcinogen.
d) Zinc:
Zinc is used in large amounts in zinc-cadmium and zinc-aluminum solders and in some base metals. When heated,
zinc oxide fumes are generated. Excessive exposure to freshly formed zinc oxide fumes can result in an illness
called metal-fume fever or "zinc chills." Symptoms include the presence of a sweetish or metallic taste in the
mouth, dryness and irritation of the throat, coughing, a feeling of weakness, fatigue, and a general malaise
condition similar to the flu.
According to Kirk-Othmer (11.5),
zinc or tin chlorides are found in some fluxes.
e) Antimony/tin:
The potential health hazard is moderate because harmful amounts of antimony or tin fumes are not generally formed.
f) Indium:
Although human exposures concerning contact with indium or its compounds have not been reported, animal studies
indicate significant lung impairment from respiratory exposures.
g) Other Metals (11.6):
Other trace metals present in base and filler metals which can give off toxic fumes include arsenic, chromium,
bismuth, cobalt, nickel, selenium, thallium, and vanadium. It should be noted that a specific PEL has not been
assigned to bismuth at this time. NIOSH has stated that arsenic is a suspected lung and lymphatic carcinogen, and
hexavalent chromium is a suspected lung carcinogen. The amount of fumes generated from these trace metals is
usually small, and hazardous concentrations are not normally found in these operations. Soldering and brazing with
filler or base metals containing these trace elements should be conducted in well-ventilated areas to
be certain that hazardous concentrations do not exist. Compounds of these metals may also be present.
1.5.4 Selection of Solder/Brazing Elements for Method Evaluation
The elements to be evaluated for analysis by ICP were selected based upon several factors which included:
- The severity of potential health hazards from soldering and brazing operations.
- Probability of occurrence of those hazards in the workplace environment.
- The instrumental considerations including compatibility of the selected analytes.
- The applicability of a dissolution procedure for the sample matrix.
For example, due to its limited use in industry, indium is not a validated element in this report. Analysis can be
performed for indium by using OSHA method no. ID-121. An
alternate ICP procedure (11.2,
11.7) is available which
can determine the elements Be, Cd, Cu, Pb, Sb, and Zn. This alternate procedure is unable to accurately quantitate
the elements Ag and Sn due to solubility problems from the H2S04/HCl digestion used.
2. Experimental Procedure
An ICP Standard Operating Procedure (SOP) is available for the ARL 3560 ICP, which details the instrumental
operating procedures used for this evaluation (11.8).
The procedure used for sample preparation has been described in OSHA Method No.
ID-206 (11.1).
Each sample taken from soldering and brazing operations is digested with HCl and HNO3 (8:1 ratio), diluted
to volume with deionized water (DI H2O) to achieve a 32% HCl/4% HNO3 mixture, and analyzed by
ICP-AES. A systematic set of experiments for the purpose of instrument evaluation was used. The experimental protocol
included:
- Investigation and correction for spectral line interferences.
- Determination of over-all analytical precision and accuracy. Analyses of 18 samples for each element (6 samples
at each test level) except for Ag where 3 samples were determined at each test level.
- Determination of analytical detection limits.
- Determination of working ranges for the elements validated.
- Evaluation of spiked Quality Control (QC) samples.
- Discussion of the digestion procedure.
- Determination of a standard reference material.
- Determination using alternate wavelengths.
- Summary.
Each of these experiments is discussed in Sections 3 through 9 below:
3. Interferences
The determination of Inter-element Correction (IEC) factors to compensate for spectral interferences is part of
the development procedure for any multi-element analytical method which uses an atomic emission spectrometer.
Spectral interferences have been minimized by the careful selection of wavelengths for the line array of the
spectrometer. Spectral interferences have also been compensated for by software which performs inter-element
corrections. This software assumes a linear relationship between analyte and interferant within the working range
limits.
3.1 Procedure
Inter-element corrections for elements likely to be found in soldering and brazing operations were experimentally
determined by identifying and then evaluating the magnitude of each interference on each analytical line. The
interferences were first qualitatively identified through peak scans and then a quantitative correction factor was
calculated. Inter-element corrections were experimentally determined by identifying and then evaluating the
magnitude of the interferences as stated below.
The identification of interferences was first determined by scanning peak profiles. Three types of peak scans
were conducted:
- the location of each analyte line peak position relative to the Mn profile line.
- the aspiration of high concentration single-element standard solutions and subsequent analysis using the
spectrometer line array to determine potential interferences.
- the re-scanning of each line during aspiration of a low concentration analyte standard solution, and multiple
higher concentrations of the interfering element to determine the extent of the interference.
The evaluated ARL ICP has a mechanical spectral scanning device (this device is called a "SAMI"
by the manufacturer) that drives the primary slit using motor steps in order to scan peak profiles over a specific wavelength range.
Scans were setup on the SAMI for ±80 motor steps to allow for sufficient resolution of the analyte peak
regardless of spectral order. This "window" was sufficient to view any interferences close enough to the analyte
peak which may be erroneously identified as the analyte. The scans were performed in 4-step increments and 2-s
signal integration time was used at each step.
Two sets of scans were analyzed in order to determine spectral interferences.
a) First spectral interference scan
These scans were used as a screening technique to determine which elements would pose a potential interference for
any given spectral line.
b) Second spectral interference scan
Information from these scans demonstrated the extent of each interference by showing the relative position of the
analyte peak to the interferant peak and the ratio of the response of the interferant to the analyte standard for
the particular spectral line.
After visually identifying potential spectral interferences from these peak profiles, the quantitative effects from
these interferences were determined by measuring the intensities of each interfering element. Apparent
concentrations resulting from these spectral line interferences on other channels were then determined. The IEC
factor was determined for each interference using the following equation:
where:
| K |
= |
IEC factor |
| B |
= |
Apparent concentration of the affected element |
| C |
= |
Concentration of the interferant (single-element) standard |
Further details for determining spectral line interferences for the ARL 3560 ICP has been described in reference
11.7.
3.2 Results
3.2.1 Peak Profiles (Motor Steps Off-profile)
The peaks for the eight element line array(s) were scanned using the "SAMI" in order to determine how close they
were to theoretical values. It has been suggested by the instrument manufacturer, as a general rule, that each
spectral line should be within ±8 motor steps from the profile peak.
[Note: For this instrument, Mn and Cd are used as profile elements. Prior to each analysis, a solution
containing the profile element(s) is scanned, the peak intensity identified, and the slit settings adjusted to give
maximum intensity for the profile element(s). Further details regarding instrument profiling can be found in
references 11.1 -
11.3, and
11.7.]
As summarized below, all the lines examined in this particular ICP-AES instrument were within ±2 motor steps.
|
Motor Steps
"off profile" |
Element |
|
| 2 |
Be |
| 0 |
Cd, Cu, Sb, Sn |
| -1 |
Ag, Pb |
| -2 |
Zn |
|
This indicates that, as a general rule, when the instrument is profiled prior to each analysis, the line
peaks of the solder elements are well within specifications.
3.2.2 Spectral Line Interferences
The interference relationships on specific wavelengths used for elements expected to be collected in
solder/brazing operations are shown in Figures 1 and 2.
Figure 1 is for those
elements affecting Pb and Sb.
Figure 2 is for those elements affecting Ag, Be, Cd, Cu, Sn, and Zn.
A chart summarizing significant spectral line interferences found during this evaluation is shown in
Table 6. This Table lists the
spectral line corrections in the order of correction (i.e. the interference of Cd on the Co channel is corrected
before the interference of Co on the Cd or Pb channel).
As can be seen from the first correction in
Table 6, an IEC correction
value of 0.00021 is necessary to compensate for the affecting element Cd on the Co channel. If Cd is present,
0.00021 µg/mL is subtracted from any Co concentration for each 1 µg/mL of Cd present in the sample. If 10 µg/mL of
Cd is present, then a 0.002 µg/mL signal is subtracted from the Co channel.
The range of IECs varied from a low of 0.000025 for Co interfering on the Cd channel to a high value of 0.0039
for Be interfering on the Sb line.
Because some solder samples may contain Bi, a study was conducted to determine the feasibility of analyzing and
correcting for Bi. An interference from Bi on Sb and Pb was noted. The instrument as set up did not contain a
channel for Bi. Bismuth is a component of some solders, especially the more recent "lead free" type. To compensate
for this, the Se channel at 196.09 nm was modified to allow a scan of the Bi line at 196.006 nm using the SAMI
scanning device. This modification to include the semiquantitative determination of Bi and any interferences from
Bi on other lines requires additional analysis time because the SAMI offsets from the Se to the Bi wavelength for
Bi measurements.
4. Precision and Accuracy
This experiment was designed to evaluate instrument performance for determining analytes normally found in a
sample matrix from soldering and brazing fume operations.
4.1 Procedure
After the IEC factors were entered into the ARL computer software, the precision and accuracy of the method were
evaluated by analyzing spiked samples. The spiking scheme for the multi-element samples was conducted in the
following manner:
The precision and accuracy for each element was evaluated by analyzing 18 spiked filter samples. Aqueous
reference standards described in Section 1.3. were used for each spike. Spike amounts were calculated for levels at
about 0.5, 1, and 2 times the OSHA TWA PEL assuming a 480-L air volume. A worst case scenario of air volumes < 400 L
was assumed for samples containing about 0.5 times the PEL for silver. Each multi-element spike was delivered to a
single mixed-cellulose ester (MCE) filter. The spikes were delivered from stock solutions using calibrated
micropipettes. A calculation error occurred regarding the mass of silver during preparation of the first
multi-element spikes. Filters spiked with silver at about 0.5 times the PEL were then prepared separately.
All spiked sample filters were digested and prepared for analysis using the procedure specified in the method (11.1).
Each sample was diluted to a 25-mL solution volume. Samples separately prepared for silver (approximately 0.5 times
the PEL) were diluted to a final volume of 10 mL. This solution volume is recommended in the method (11.1)
to provide increased sensitivity for silver.
The instrument was calibrated as stated in Section 1.4. These samples were determined using a two-point
calibration consisting of standards listed in
Table 3 and a reagent blank.
4.2 Results
All sample results were examined in terms of precision (CV) and amount of error. Analytical error (AE) for each
element is calculated as:
| ± AE % = 100 [|Mean Bias| + 2(CV)] 95% confidence |
From the summary in
Table 7, it can be seen that
the AE ranged from 2.3% for antimony to approximately 17% for silver and lead. The precision range (CV) was from
0.01 for many of the analytes to 0.06 for silver and lead. The bias varied from -0.011 for beryllium to +0.055 for
zinc.
5. Detection limit, Background Equivalent Concentration, and Short-Term Precision
5.1 Procedure
The procedure for the determination of detection limits, Background Equivalent Concentration (BEC) and short-term
reproducibility for the ARL 3560 ICP has been previously described (11.7).
The detection limits in this evaluation were determined after aspirating multielement calibration standards for a
5-s integration time. An 11 exposure sequence was used for the blanks and 10 exposures for the standards.
Multi-element calibration standards (STND SOLNS 1 and 2) shown within
Table 3 were used. The
manufacturer's software algorithms calculate the qualitative detection limit (DL) as two times the standard
deviation.
The qualitative detection limit for the ARL instrument is calculated as:
| DL = (A × SDI × C)/(I - Io) |
Where:
| A |
= |
2 (qualitative detection limit) or |
| A |
= |
10 (quantitative detection limit) |
| SDI |
= |
Standard Deviation of the Background Intensity |
| I |
= |
Total Intensity |
| C |
= |
Analyte Concentration, the concentration of the calibration standard |
| Io |
= |
Background Intensity (determined from reagent blank) |
The BEC is defined as the concentration of an analyte that is equal to the net intensity of the background signal
for that analyte:
Where:
m = (C)/(I - Io) = slope of the calibration curve
Note: The manufacturer's software algorithms automatically calculate the qualitative detection limit using two times
the standard deviation (2SD) of the blank, among other considerations. Although the OSHA Inorganic Method Validation
Protocol (11.9) states a
qualitative limit shall be determined with 3SDBL in the calculations 2SDBL is accepted here to
allow for future performance comparisons using the same instrument and software.
5.2 Results
The results for short-term instrument precision are reported in
Table 8 as the Coefficient of
Variation (CV) and Background Equivalent Concentration (BEC). At concentration levels equal to or greater than the
BEC, short-term precision ranges are normally approximately 0.5% to 1.5% (as CV), provided linearity of the spectral
response function is maintained. Short-term precision should be 1% or better for simultaneous systems (11.10).
As can be seen in the last column of
Table 8, the CV is < 1% for
all lines. The qualitative detection limits varied from 0.001 µg/mL for beryllium to 0.05 µg/mL for lead. The BEC
values ranged from 0.02 µg/mL for beryllium to 1.9 µg/mL for lead.
Detection limits determined approximately one year after these results were obtained show improved performance
for most elements:
|
| ELEMENT |
QUAL DL |
|
Ag
Be
Cd
Cu
Pb
Sb
Sn
Zn |
0.0043
0.0002
0.0054
0.0055
0.0425
0.0481
0.0396
0.0031 |
|
Detection limits are very dependent on maintenance and operating conditions, and should be periodically
checked to assess instrument performance and the need for maintenance.
6. Working Range
The calibration used for routine analysis of solder samples is a first order regression. For the calibrations, the
ARL 3560 ICP computer software calculates a linear regression equation for each element from the intensities (counts)
of two measurements (a reagent blank and a reference solution usually in the range of 1 to 10 µg/mL). An evaluation
to determine the appropriate lower and upper concentration ranges for each validated or screening element was
performed using these calibrations.
6.1. Procedure
The linearity was evaluated using a procedure previously described (11.7).
The linearity of the calibration curves was checked by analyzing several standard solutions within the range from 5
to 10 times the calculated detection limit up to 1,000 µg/mL. The upper range was limited either by reaching a
detector saturation level or by exceeding the value of the highest standard stock solution used (1,000 µg/mL).
6.2. Results
The highest concentration of most of the standards used was 1,000 µg/mL. For a few elements, a 1,000 µg/mL
concentration did not saturate the detector nor was the linear range exceeded. Standard solutions exceeding a
concentration of 1,000 µg/mL were available for these elements and linearity determinations beyond 1,000 µg/mL were
performed. These exceptions were: aluminum, magnesium, lead, bismuth, and antimony.
The upper working ranges (µg/mL) and the concentrations at which the photomultiplier tubes (PMTs) become
saturated for elements at the wavelengths designated in the array are summarized in
Table 9. The PMTs should
become saturated at a value between the Upper and Saturation concentrations reported in the Range column.
In Table 8, column five
contains the lower quantitative detection (LQD) limit and column six the upper range of linearity for the brazing
and solder elements. For the ARL instrument, the LQD is five times the qualitative detection limit. The quantitative
detection limits ranged from 0.005 µg/mL for Be to 0.2 µg/mL for Pb and Sb.
The linear range was evaluated using a 5-s integration time. The upper concentration limit for the validated
elements ranged from 20 µg/mL for Be to 1,000 µg/mL for Pb and Sb. The optimum working range for most elements
exceeded 100 µg/mL.
7. Evaluation of Spiked Quality Control (QC) Samples
7.1. Procedure
Sets of QCs were determined for a final evaluation of the precision and accuracy of the procedure. These QCs were
independently prepared on spiked filters. The concentration levels for some of the analytes on these QCs were lower
than those in the previous precision and accuracy evaluation. With the exception of beryllium, the concentrations
for the prepared spikes were based upon calculations using a 4-h sampling period and an air volume of 480 L.
The concentrations ranged from 0.1 to 1 times the TWA PELs as follows:
|
| ELEMENT |
× PEL |
|
Ag
Be
Cd
Cu
Pb
Sb
Sn
Zn |
0.5
(Ceiling or STEL)
0.1
0.5 to 1
1
0.5 to 1
0.5 to 1
1 |
|
To determine if precision and accuracy could be improved using smaller solution volumes, all samples were
diluted to a final solution volume of 10 mL.
7.2. Results
The data, as summarized from
Table 10, provide the following method performance information:
With the exception of Be in one QC set, the AE was less than 25%. The Precision (CV) was better than 0.14 for all
samples. With the exception of Cu in the last QC set, the Mean varied within ±10%. For the most part, these results
are similar or better than those presented in Section 4, and indicate a 10 mL solution volume can be used as an
alternative dilution. It is unknown why one set of beryllium samples was ± 35% AE. Samples to be analyzed for
beryllium at Ceiling or STEL levels should be diluted to 10 mL to improve sensitivity.
8. Digestion Procedure Discussion
Experiments were previously conducted (11.1,
11.11) to evaluate the
digestion procedure for solders. Following validation of the method, spiked samples of the eight elements were
prepared by an independent group within the OSHA SLTC and had been routinely analyzed using the JY-70 ICP. Some of
the spiked samples had low antimony recoveries.
As shown by the precision and accuracy, and independent QC results within this report, the recoveries for antimony
are adequate if the digestion procedure in OSHA method no.
ID-206 is followed. The loss of antimony was attributed to incomplete wetting of the filters with HCl before
addition of HNO3.
9. Determination of a Standard Reference Material
A Standard Reference Material (SRM) of solder from National Institute of Standards and Technology (NIST),
containing certified values as reported below, was determined as "blind samples".
NIST SRM 127b Solder
(40 Sn-60 Pb)
|
| Element |
Percent by Weight |
|
| Tin |
39.3 |
| Antimony |
0.43 |
| Arsenic |
.01 |
| Bismuth |
.06 |
| Copper |
.011 |
| Nickel |
.012 |
| Silver |
.01 |
| Lead |
60.17* |
|
| * actual amount is not listed on the certificate. The percent was calculated by
subtracting the reported elements from 100% and assuming the remainder as lead. |
9.1. Procedure
Three "blind samples" containing the SRM were routinely analyzed by a laboratory chemist. Sample preparation for
bulk material was carried out according to OSHA method no.
ID-206. Additional details are described below:
|
sample wt
range (mg) |
sample wt
median (mg) |
final sample
vol (mL) |
sample
(mg/mL) |
|
| 10-14 |
12 |
250 |
0.048 |
|
9.2. Results
Due to the weights of SRM used, the values found for Ag were below both the quantitative and qualitative
detection limits. The values for Cu, Sb, and Zn were below the quantitative but above the qualitative detection
limits. Therefore, the values for Cu, Sb, and Zn could not be used and merely indicate that these elements were
present in the sample matrix.
Although Zn was found in the SRM, it was not reported on the accompanying certificate. It is possible that due to
the large amounts of Pb and Sn present in the bulk material these elements might possibly give an apparent value for
Zn, as the spectral corrections for this method are for low concentrations typically found in workplace atmospheres.
Caution must be observed when using this method as a screening tool to identify unknown elements in a bulk matrix.
Minor elemental quantities found for elements in bulks must he considered as apparent values only, and may have to
be confirmed by other methods such as Atomic Absorption Spectroscopy.
From the results reported below it is evident that if the minor components are to be reported, two sample
aliquots will have to be taken; one for the more concentrated aliquot used for quantitating the values for Sb, Cu,
and Ni, the other a larger weight for minor components.
A suggested scheme for solder bulk materials is to weigh a 10 to 20 mg aliquot; dilute to 250 mL to obtain the
major components within the linear working range; and prepare another sample using approximately a 100-mg sample
weight and 100-mL sample solution volume. This will enable both major and minor components to be identified,
provided both sample results are carefully scrutinized for exceeding upper linear range limits, detection limits, or
interferences potentially existing in a variable sample matrix.
|
| |
------------------Wt %-------------------- |
| |
Pb |
Sn |
Sb |
Cu |
Ni |
Ag |
|
| Reference values |
60.17 |
39.3 |
0.43 |
0.011 |
0.012 |
0.01 |
| Avg Recovery |
60.4 |
38.1 |
n.d. |
n.d. |
n.d. |
n.d. |
| F/T |
100. |
96.9 |
- |
- |
- |
- |
|
| n.d. = none detected |
10. Determinations using Alternate Wavelengths
10.1. Introduction
An ideal situation for emission spectroscopy would be to have many alternate spectral lines available for
analysis of each element to assist in characterizing the sample and minimize error. Unfortunately, there is a finite
amount of physical space available for the installation of spectral lines in the array of a simultaneous direct
reading emission spectrometer. This limits the number of element channels that may be installed. An alternative to
increasing channels for the ARL 3560 ICP is to "create" lines using the SAMI scanning mechanism. The SAMI, as
previously described in Section 3.1., is a stepper-motor-controller which allows for mechanical profiling of the
slit image and the relative wavelength position. The SAMI is also routinely used for background corrections, moving
off and on a fixed wavelength at specified intervals. This ability to move away from the fixed line position allows
for examination of other emission lines nearby. These emission lines, referred to as "free SAMI" lines, are limited
because they must be in close proximity to element channels previously installed by the instrument manufacturer.
10.2. Procedure
Using the scanning ability of the SAMI mechanism, free SAMI lines (optional wavelengths) for the elements Bi, Cd,
Be, and As were programmed into the manufacturer's software as described in the table below:
Selected Free SAMI Lines
|
| Ele(2) |
Wavelength(2)
(nm) |
Ele(1) |
Wavelength(1)
(nm) |
Offset
(Steps) |
Spectral
Order |
|
Bi
Cd1
Cd3
Be3
As1 |
196.006
228.801
214.433
217.506
228.812 |
Se
Co
Te
Sb
Co |
196.090
228.616
214.275
217.581
228.616 |
218.00
-320.00
-274.00
192.00
-338.00 |
3
2
2
3
2 |
|
Where:
Ele(1) = original channel modified to allow a scan of Ele(2).
For example, the Co 228.616 nm line [Ele(1)] was modified by the SAMI (offset -320.00 divisions) to scan the
228.801 nm line for Cd [Ele(2)]. |
|
Detection limits and BECs were then determined for these free SAMI lines using the procedure discussed in
Section 5.
10.3. Results
The following BEC and detection limit (DL) values were calculated after determinations using the concentration of
standards specified (Std Concn). Reagent blanks were prepared in DI H2O. The DL was reported as a
qualitative detection limit (A=2, as stated in Section 5.1).
BEC/DL - Free SAMI and Original Spectral Lines
|
Channel
Name |
Blank
Int. |
Blank
SD |
Std
Int. |
Std
Concn |
BEC |
DL |
Wavelength
(nm) |
|
| As |
5.8799 |
0.2734 |
143.98 |
100.00 |
4.2578 |
0.3960 |
189.042 |
| As1 |
6.1419 |
0.1108 |
67.950 |
100.00 |
9.9371 |
0.3586 |
228.812 |
| As |
2.7624 |
0.1230 |
1,369.1 |
1,000.0 |
2.0217 |
0.1801 |
|
| As1 |
4.7699 |
0.0612 |
625.87 |
1,000.0 |
7.6798 |
0.1971 |
|
|
| Be |
4.5310 |
0.2074 |
1,450.4 |
10.000 |
0.0313 |
0.0029 |
313.042 |
| Be3 |
4.3889 |
0.0232 |
29.060 |
10.000 |
1.7790 |
0.0188 |
217.506 |
| Be |
4.5310 |
0.2074 |
|
100.00 |
saturated detector |
| Be3 |
4.3889 |
0.0232 |
246.92 |
100.00 |
1.8096 |
0.0191 |
|
|
| Bi |
3.3873 |
0.0172 |
7.1453 |
100.00 |
90.134 |
0.9166 |
196.006 |
| Bi |
3.3527 |
0.0302 |
11.011 |
200.00 |
87.555 |
1.5754 |
|
|
| Cd1 |
4.2365 |
0.0168 |
203.77 |
10.000 |
0.2123 |
0.0017 |
228.801 |
| Cd3 |
3.5647 |
0.0164 |
314.33 |
10.000 |
0.1147 |
0.0011 |
214.433 |
| Cd |
1.7259 |
0.0164 |
98.449 |
10.000 |
0.1784 |
0.0034 |
226.502 |
| Cd1 |
4.2365 |
0.0168 |
1,000.0 |
|
saturated detector |
| Cd3 |
3.5647 |
0.0164 |
|
1,000.0 |
saturated detector |
| Cd |
1.7259 |
0.0164 |
|
1,000.0 |
saturated detector |
|
The elements (Channel Name) which are not followed by a number are the original
installed lines.
Blank Int. and Std Int. are blank and standard intensities in K-pulses (this is a unique term used by the
manufacturer to designate intensity counts).
All concentration (Concn) values (Std, Concn, BEC, and DL) are in µg/mL. |
These additional lines may be used as elemental confirmation to avoid the reporting of "false positives"
which can occur at concentrations near the detection limit due to matrix effects, uncharacterized spectral, or other
interferences. The application of these alternate lines will extend the analytical time taken for each sample, and
subsequently affect the amount of sample volume used during sample preparation. Sample solution volumes of 10-mL may
be insufficient if numerous additional lines are scanned and more than one sample determination is necessary. Some
notes regarding the sensitivity of each line are listed below:
Bismuth:
The Bi channel is relatively insensitive, yielding high BEC and DL values at concentrations of from 100 to 200
µg/mL. However, as previously discussed in Section 3, this line may be used as confirmation for the presence of Bi
in "lead free" bulk solder samples.
Cadmium:
The alternative lines for Cd have improved detection limits of a magnitude of about three times that of the Cd
226.502 nm line. This increased sensitivity is important as the PEL for Cd has recently been lowered significantly
(29 CFR 1910.1027 or 29 CFR 1926.63). It is important to note there is a major As spectral interference on the
alternate Cd 228.801 nm line.
Arsenic:
Although at first glance it appears that the alternate As 228.812 line has a comparable detection limit with the As
189.042 line, this is misleading. Upon further examination, intensities from the same sample concentrations are
about two times greater for the As 189.042 line. The reason for the similar detection limits results from the
imprecision of the As 189.042 blank (being about two times higher than that of the As 228.512 line). In addition,
there is a major spectral interference on the alternate As line resulting from cadmium.
Beryllium:
The original Be 313.042 line is approximately ten times more sensitive than the alternate Be 217.506 line; however,
the 217.506 line has a higher upper range and reduces the requirement of further sample dilution for samples
containing high concentrations of Be.
Summary
In conclusion, this evaluation demonstrated that the ARL ICP instrument adequately determined the eight elements
commonly found in soldering and brazing fumes within the ranges specified. Compliance with OSHA Transitional or Final
Rule PELs for solders can be determined using this instrumental procedure (Note: At the time of this writing the
OSHA Final Rule Limits were stayed). Precision and accuracy, detection limits, and linear working ranges were all
acceptable for air samples taken near the PELs for the eight elements validated. As previously stated in Section 7.2.,
sample dilutions using 10 mL solution volumes should result in improved sensitivity for Be. Sensitivity for other
elements will also improve when using the 10-mL final solution volume; however, a drawback is the amount of sample
available for repeat analysis. Due to the design of the automatic sampler/nebulizer sample introduction system, about
5 to 6 mL of sample is needed for each analysis.
A drawback to this method is the acid digestion and dilution media; the acid matrix used (32% HCl/4% HNO3)
is very corrosive, and care must be exercised when handling solutions. As discussed in a previous work (11.1,
11.11), care must be taken
during the digestion to prevent the loss of Sb or Sn. Special attention must be taken regarding the order of addition
of the digestion acids. After adding the HCl, it is recommended to wait 5 min before adding HNO3.
This analytical method should be applicable with minor modifications to any ICP containing analytical lines for the
eight elements evaluated. Further work needs to be conducted to evaluate the ability of this ICP for analyzing Cd at
the new PEL of 0.005 mg/m³.
Addendum
During this and a previous evaluation, the existence of a few "bugs" were noted in the software provided by ARL.
Although this software (and computer hardware) is no longer available, current users should be made aware of the
problems. One problem occurs during repeated calibration of the instrument. During each calibration, an
inter-element correction is applied if two interfering elements are present in the same calibration standard
solution. For a few of the elements having interferences, the respective inter-element correction being
applied became additive with the next calibration (i.e. if 10 µg/mL Sb was present in a calibration solution
containing 10 µg/mL Zn, an inter-element correction of 0.03 was applied such that the Sb concentration
became 10.03 µg/mL to compensate for the additive effect of Zn on the Sb line. The software would not erase this 10.03
µg/mL calibration concentration and the next calibration valid result in Sb being recognized as 10.06 µg/mL, even
though there was only 10 µg/mL Sb in the solution). The correction would continue to be additive with each calibration
unless the user applied a software program called CSET to "reset" the concentrations to their original values after
each calibration. This was somewhat time-consuming to perform and was not widely known by the manufacturer's personnel
nor was it documented in any manuals.
For other elements, the concentration would be "reset" to the original concentration for the next calibration. Most
of the elements which had an interference present in the same calibration solution were not affected.
11. References
11.1 Occupational Safety and Health Administration Salt Lake Technical Center: ICP
Analysis of Metal/Metalloid Particulates from Solder Operations by D.C. Cook (USDOL/OSHA-SLTC Method No.
ID-206). In OSHA Analytical Methods Manual. 2nd ed.
Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1991.
11.2 Occupational Safety and Health Administration Salt Lake Technical Center:
Metal and Metalloid Particulates in Workplace Atmospheres (ICP Analysis) by J. Septon (USDOL/OSHA Method No.
ID-125G). In OSHA
Analytical Methods Manual. 2nd ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists,
1991.
11.3 Occupational Safety and Health Administration Salt Lake Technical Center:
Metal and Metalloid Particulates in Workplace Atmospheres (ICP Analysis) (Backup Data Report) by J. Septon
(USDOL/OSHA Method No. ID-125G).
In OSHA Analytical Methods Manual. 2nd ed. Cincinnati, OH: American Conference of Governmental Industrial
Hygienists, 1991.
11.4 "Air Contaminants; Final Rule": Federal Register 54:12 (19 Jan. 1989). pp.
2923-2960; also 54:127 (5 July 1989). pp. 28054-28061.
11.5 Solders & Brazing Alloys. In Kirk-Othmer Encyclopedia of Chemical Technology,
3rd Ed., Vol. 21, edited by H.F. Mark, D.F. Othmer, C.G. Overberger, and G.T. Seaborg. New York, NY: John Wiley &
Sons, 1983.
11.6 National Institute for Occupational Safety and Health: Health and Safety
Aspects of Soldering and Brazing [DHEW(NIOSH) Pub. No. 78-197]. Cincinnati, OH: Division of Technical Services,
September, 1978.
11.7 Occupational Safety and Health Administration Salt Lake Technical Center:
Welding Fumes ICP Backup Data Report (ARL 3560) by J. Septon. Salt Lake City, UT, 1991.
11.8 Occupational Safety and Health Administration Salt Lake Technical Center: ICP
Analysis Reference Guide, OSHA Lab (Comprehensive Version) by J. Septon. Salt Lake City, UT, 1991 (unpublished).
11.9 Occupational Safety and Health Administration Salt Lake Technical Center:
Evaluation Guidelines of the Inorganic Methods Branch In OSHA Analytical Methods Manual. 2nd ed.
Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1991.
11.10 Arellano, S.D., M.W. Routh, and P.D. Dalager: Criteria for evaluation of
ICP-AES performance. Amer. Lab.: 20-32 (August, 1985).
11.11 Occupational Safety and Health Administration Salt Lake Technical Center:
Recovery of Antimony and Other Elements Using the Digestion Procedure for Solders by R. Douglas and D.C. Cook.
Salt Lake City, UT, 1991. (unpublished)
Table 1
Specifications for ARL 3560 Simultaneous AES-ICP
|
R.F. GENERATOR
Generator Model # [kW] Henry
Incident Power [W]
Reflected RF Power [W]
P.A. Plate Supply Voltage [V]
P.A. Plate Current [mA]
P.A. Grid Current [mA]
P.A. Filament Voltage [V]
P.A. Tune Setting
P.A. Load Setting
(where P.A. = Power Amplifier)
EXCITATION
Plasma Observation Height [mm]
Coolant Gas Flow [L/min]
Coolant Gas Pressure [psi]
Plasma Gas Flow [L/min]
Plasma Gas Pressure [psi]
Carrier Gas Flow [L/min]
Carrier Gas Pressure [psi]
Snout Gas Flow [L/min]
Nebulizer Uptake Rate [mL/min]
Nebulizer [type]
Peristaltic Pump Used? [yes/no]
Pre-Integration Flow Time [s]
Integration Time [s]
SPECTROMETER
Instrument Model
Grating [lines/mm]
Primary Slit Size [µm]
Profiling Element
Profile Point [peak dial div.]
Vacuum [µm]
Path Length [m] |
2.5
1175
0
4100
700
125
7.5
190
050
15
12
25.5
0.8
21.5
1
30.5
1.5
2 to 3
Meinhard-Type C
no
30
5
3560
1080
20
Mn
496*
20
1 |
| * For the SAMI, 100 dial divisions = 400 motor steps |
COMPUTER CONFIGURATION
Disk Drive Capacity
Computer System
Software: |
30 MEG fixed and 1.2 MEG floppy
Micro 11/53 DEC Computer
DPS/TSX+ |
|
Table 2
Line Library for OSHA ARL 3560 Simultaneous AES-ICP
|
| Element |
Wavelength
(nm) |
Spectral Order |
|
| |
Ag
Al*
As*
Au
B
Be
Bi*
Ca*
Cd
Co*
Cr*
Cu
Fe1*
Fe2*
Ga
Ge
In
Mg*
Mn*
Mo*
Ni*
Os
Pb
Pd
Pt
Rh
Sb
Se*
Si*
Sn
Te
Tl
U
V*
W
Zn
Zr |
328.068
308.215
189.042
242.795
249.680
313.042
196.006
393.366
226.502
228.616
267.720
324.754
259.940
271.440
417.205
209.430
230.606
279.080
257.610
202.030
231.604
225.585
220.353
360.955
265.945
343.489
217.581
196.090
288.158
189.980
214.275
190.864
409.014
310.230
239.709
213.856
343.823 |
2
2
3
3
3
2
3
2
3
2
3
2
3
3
1
3
3
2
3
3
3
3
3
2
2
2
3
3
2
3
2
3
1
2
3
3
1 |
|
| Measurement of bolded elements are validated in this report. Elements
with an asterisk (*) are screened. Bismuth (Bi) is also in the line library through modification of the Se channel. |
Table 3
Calibration Standards
|
| NO |
ELE |
STD SOLN |
CONCN |
|
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22 |
Ag
Be
Cd
Pb
Cu
Sb
Sn
Zn
Al
As
Ca
Co
Cr
Fe
Mg
Mn
Mo
Ni
Se
Si
V
Bi |
1
1
1
1
2
2
2
2
3
3
3
3
3
3
3
3
3
3
3
3
3
3 |
1.
1.
10.
10.
1.
10.
10.
10.
10.
10.
10.
10.
10.
100.
10.
10.
10.
10.
10.
10.
10.
100. |
|
|
Where:
ELE
STD SOLN
CONCN |
Element
Number of Calibration standard
Concentration (µg/mL) of calibration standard |
|
STD SOLN 1 - was prepared in an amber-colored glass bottle to protect the Ag from photo-decomposition.
STD SOLN 3 - This standard is used only for calibration of the screened elements. The digestion and analysis are not
optimized for these elements.
Calibration is accomplished using a two-point calibration curve with the concentration for each element listed
above. A reagent blank was used as the low standard. Each element calibrated is contained in one of three separate
calibration standards (STD SOLN). For example, STD SOLN 1 contains Ag, Be, Cd, and Pb.
All solutions were prepared in a 32% HCl/4% HNO3 acid mixture including the reagent blank.
Table 4
Air Contaminants - OSHA Permissible Exposure Limits1
|
| Element |
Substance Exposed to |
Transitional PEL
(mg/m³) |
Final Rule PEL
(mg/m³) |
| |
|
TWA |
CEILING |
TWA |
STEL |
CEILING |
|
| Ag |
Metal dust and fume (as Ag) |
0.01 |
|
0.01 |
|
|
| Be2 |
Be and compounds (as Be) |
0.002 |
0.005³ |
0.002 |
0.005³ |
0.025 |
| Cd |
Fume
Dust |
0.1
0.2 |
0.3
0.6 |
0.1
0.2 |
0.3
0.6 |
|
| Cu |
Fume (as Cu)
Dusts amd mists (as Cu) |
0.1
1 |
|
0.1
1 |
|
|
| Pb |
Inorganic |
(see 29 CFR 1910.1025, PEL is 0.05 mg/m³) |
| Sb |
Sb and compounds (as Sb) |
0.5 |
|
|
0.5 |
|
| Sn |
Inorganic compounds
except oxides (as Sn)
Tin oxide (as Sn) |
2
|
|
2
2 |
|
|
| Zn |
Zinc chloride fume
Zinc oxide fume |
1
5 |
|
1
5 |
2
10 |
|
| Note: underlined values are those TWA PELs selected for validation. |
|
STEL duration is for 15 min unless otherwise noted.
1 From reference 11.4
2 Beryllium also has Transitional Peak PEL concentration limit of 0.025 mg/m³.
3 Both the Transitional Ceiling Limit and the STEL for beryllium are
for a maximum 30-min duration. |
|
Note: This validation was performed using the Final Rule Limits bolded above. With a
couple of exceptions the Final Rule Limits for the elements or compounds listed are identical to the Transitional
Limits. Analysis of samples by the ARL 3560 using OSHA method no.
ID-206 should perform well regardless of whether
Transitional or Final Rule Limits are used. After this validation was performed, the Final Rule Limits were vacated.
For determination of exposures to rosin core solder pyrolysis products, please see OSHA method no. 54 for
formaldehyde. |
|
Table 5
Common Solder Alloys*
|
| ---------Composition %--------- |
Melting
range, °C |
Use |
| Sn |
Pb |
Cd |
Bi |
Ag |
Sb |
|
|
|
| 63 |
37 |
- |
- |
- |
- |
183 |
eutectic solder for electronic application |
| 60 |
40 |
- |
- |
- |
- |
183-190 |
high quality solder |
| 50 |
50 |
- |
- |
- |
- |
183-216 |
general-purpose solder, plumbing |
| 40 |
60 |
- |
- |
- |
- |
183-238 |
wiping solder, radiator solder |
| 30 |
70 |
- |
- |
- |
- |
183-255 |
machine and torch soldering |
| 20 |
80 |
- |
- |
- |
- |
183-277 |
automotive-body solder |
| 95 |
- |
- |
- |
- |
5 |
235-240 |
refrigeration soldering |
| x |
0 |
0 |
x |
x |
0 |
|
"Pb free" plumbing solderinga |
| 62 |
36 |
- |
- |
2 |
- |
179 |
soldering silver surfaces |
| 1 |
97.5 |
- |
- |
1.5 |
- |
309 |
high temperature soldering |
| 15.5 |
32 |
- |
52.5 |
- |
- |
90 |
fusible links |
| 13 |
27 |
10 |
50 |
- |
- |
70 |
low melting solder |
|
* Modified from reference
11.5
- not present or reported
a This solder also contains copper, and is patented alloy containing unspecified amounts of elements
designated with an x. These were confirmed at the SLTC by ICP and XRF analysis. A sample of "Safe Flo (TM) Silver"
solder, gauge 0.084, from Oatey, Cleveland, OH 44135 was used for this confirmation.
Note: Because of recent requirements of the Safe Drinking Water Act and certain state laws restricting lead-solder
has been developed that contains no lead, antimony, or cadmium. These "lead free" solders have become popular in the
marketplace. |
Table 6
Interferences
|
| |
Channel |
Int. |
IEC factor |
|
|
Co
V
Cd
Cd
Sn
Sn
Sn
Cu
Ag
Ag
Pb
Pb
Pb
Pb
Sb
Sb
Sb
Sb
Zn
Be
Sb
Sb
Zn
Ag
Sn
Sb
Sb
Pb
Bi*
Bi*
Bi* |
Cd
Be
Ni
Co
Al
V
Mo
Mo
Mn
V
Ni
Co
Al
Mo
Ni
Mo
Fe1
Al
V
Ni
V
Zn
Pb
Cu
Cu
Be
Be
Bi*
Bi*
Be
Fe
Al |
0.00021
0.00093
0.00011
0.000025
0.00040
0.0016
0.00013
0.00061
0.00012
0.00016
0.00074
0.00044
0.0010
0.000235
0.0015
0.00035
0.00022
0.0012
0.0014
0.0038
0.00046
0.0015
0.00081
0.0013
0.000062
0.00042
0.0039
0.0003
0.001
0.009
0.37
0.049 |
|
Int.
Channel |
= Interference
= The Affected Channel |
| * Bi line available through modified Se channel. |
|
This table is organized in the following manner:
The first two columns list two elements. The first element is the channel that is affected by the second element
(the affecting element). The inter-element Correction Factor (IEC) for the affected element is shown in
the third column.
Table 7
Precision and Accuracy
|
| |
Element |
|
|
|
|
|
Ag |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
2.21 to 10.1
+0.036
0.066
±16.9% |
|
Be |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
0.5 to 2.0
-0.011
0.015
±4.0% |
|
Cd |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
24.0 to 100.
+0.054
0.026
±10.5% |
|
Cu |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
24.0 to 100.
+0.017
0.039
±9.5% |
|
Pb |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
12.0 to 48.0
+0.050
0.058
±16.7% |
|
Sb |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
120. to 480.
+0.008
0.007
±2.3% |
|
Sn |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
480. to 1,920.
+0.033
0.015
±6.3% |
|
Zn |
Range taken (µg)
Bias
CV1 (Pooled)
Analytical Error (Total) |
=
=
=
= |
600. to 4,800.
+0.055
0.013
±8.0% |
|
| Note: Analytical errors (AE) of < ±25% are acceptable for all elements validated
except lead. [Lead is required to have < ± 20%, see 29 CFR 1910.1025(d)(9) for further information] |
Table 8
Detection Limits, BEC, Ranges, and Short-Term Precision
|
| ELEMENT |
WAVELENGTH |
*QUAL |
----*RANGE---- |
|
|
(nm) |
*BEC |
DL |
LOWER |
UPPER |
CV |
|
Ag
Bi
Be
Cd
Cu
Pb
Sb
Sn
Zn |
328.068
196.006
313.042
226.502
324.754
220.353
217.581
189.980
213.856 |
0.4857
90.
0.0177
0.1645
0.4982
1.8816
1.5882
0.6104
0.1897 |
0.0109
1.0
0.0010
0.0131
0.0142
0.0470
0.0388
0.0139
0.0241 |
0.0545
5.0
0.0050
0.0655
0.071
0.235
0.194
0.0695
0.1205 |
500
1,000
20
250
250
1,000
1,000
900
500 |
0.0035
0.03
0.0045
0.0031
0.0019
0.0066
0.0092
0.0015
0.0028 |
|
Results were obtained from determinations performed on 21-Jan-92
| * |
= |
BEC, DL, and RANGE values reported as µg/mL |
| DL |
= |
Qualitative Detection Limit (SD = 2) |
| BEC |
= |
Background Equivalent Concentration |
| CV |
= |
Short-Term Precision (based on three exposures) |
| LOWER |
= |
Quantitative Detection Limit (5 × DL) |
| UPPER |
= |
Upper Working Range |
| Integ. time |
= |
5 s |
|
Table 9
Upper Range and Saturation Concentrations
|
|
|
|
-----Range----- |
Element
Name |
Wavelength
(nm) |
Order |
Upper
> |
Saturation
< |
|
Ag*
Al
As
Be*
Bi
Ca
Cd*
Co
Cr
Cu*
Fe1
Fe2
Mg
Mn
Ni
Pb*
Sb*
Se
Si
Sn*
V
Zn* |
328.068
308.215
189.042
313.042
196.006
393.366
226.502
228.616
267.720
324.754
259.940
271.440
279.080
257.610
231.604
220.353
217.581
196.090
288.158
189.980
310.230
213.856 |
2
2
3
2
3
2
3
2
3
2
3
3
2
3
3
3
3
3
2
3
2
3 |
500
5,000
700
20
10,000
15
250
300
300
250
200
400
3,000
60
600
1,000
1,000
1,000
1,000
900
1,000
500 |
|
1,000
10,000
50
10,000
500
500
500
500
5,000
100
1,000
3,000
3,000
1,000
1,000
1,000 |
|
|
Range is listed as µg/mL
* Validated elements |
Table 10
Quality Control Samples
|
|
N |
MEAN |
STD DEV |
CV |
AE
(±%) |
µg TAKEN
---(range)--- |
|
QC Set 1
|
|
Ag
Be
Cd
Cu
Pb
Sb
Sn
Zn |
3
3
3
3
3
3
3
3 |
0.930
0.962
0.935
1.089
0.939
0.898
0.917
0.937 |
0.023
0.015
0.021
0.027
0.024
0.047
0.023
0.011 |
0.025
0.015
0.023
0.025
0.026
0.052
0.025
0.012 |
11.9
6.9
11.1
14.0
11.2
20.6
13.3
8.7 |
|
2.41
0.201
5.02
25.2
20.1
126.
504.
315. |
3.66
0.305
7.62
47.4
30.5
237.0
948.
592. |
|
QC Set 2
|
|
Ag
Be
Cd
Cu
Pb
Sb
Sn
Zn |
3
3
3
3
3
3
3
3 |
0.998
1.091
0.998
0.993
1.023
0.933
0.964
0.976 |
0.013
0.141
0.010
0.011
0.021
0.023
0.012
0.007 |
0.013
0.130
0.010
0.011
0.020
0.024
0.013
0.007 |
2.9
35.0
2.3
3.0
6.4
11.5
6.1
3.9 |
|
2.77
0.228
5.77
33.
23.1
165.
660.
412.5 |
3.46
0.267
7.20
51.
28.8
258.
1032.
645. |
|
QC Set 3
|
|
Ag
Be
Cd
Cu
Pb
Sb
Sn
Zn |
3
3
3
3
3
3
3
3 |
0.927
0.934
0.898
1.139
0.930
0.901
0.894
0.912 |
0.005
0.039
0.043
0.029
0.035
0.025
0.023
0.029 |
0.006
0.041
0.048
0.026
0.038
0.027
0.025
0.032 |
8.5
14.9
19.8
19.0
14.6
15.3
15.7
15.2 |
|
2.60
0.217
5.42
34.8
21.7
174.
696.
435. |
3.43
0.286
7.15
54.0
28.6
270.
1080.
675. |
|
| AE = Analytical Error |
SPECTRAL LINE INTERFERENCES
Figure 1: Interactions of Pb and Sb |
SPECTRAL LINE INTERFERENCES
Figure 2: Interactions of Ag, Be, Cd, Cu, Sn, Bi and Zn |
i.e. as shown from Figure 2, Be has an interference from V; Be interferes with V, Sn, Bi and Al.
|
|